Case report


Partially Covered Transjugular Intrahepatic Portosystemic Shunt Endograft for Repair of a Primary Aortoenteric Fistula

Akshaar Brahmbhatt MD1Anisha Shetty BS2Salman Mirza DO3
1Department of Imaging Sciences, University of Rochester Medical Center, 601 Elmwood Ave, Box 648, Rochester, 14642, New York, United States, 2Department of School of Medicine, University of Rochester Medical Center, 601 Elmwood Ave, Rochester, 14642, New York, United States, 3Department of Vascular and Interventional Radiology, Geisinger Community Medical Center, 1800 Mulberry Street, Scranton, 18510, Pennsylvania, United States
Date of Submission: 16-Oct-2018, Date of Acceptance: 29-Nov-2018, Date of Web Publication: 10-Dec-2018.
Corresponding Author:
Corresponding Author

Anisha Shetty BS

Department of School of Medicine, University of Rochester Medical Center, 601 Elmwood Ave, Rochester, 14642, New York, United States
E-mail: anisha_shetty@urmc.rochester.edu

Corresponding Author:
Corresponding Author

Anisha Shetty BS

Department of School of Medicine, University of Rochester Medical Center, 601 Elmwood Ave, Rochester, 14642, New York, United States
E-mail: anisha_shetty@urmc.rochester.edu

DOI: 10.25259/AJIR-45-2018 Facebook Twitter Google Linkedin

Aortoenteric fistulae (AEF) are a rare cause of life-threatening gastrointestinal hemorrhage. AEF are divided into primary and secondary types. Secondary fistulae are more common and occur after prior aortic interventions, namely aneurysm repairs.Less common is primary aortoenteric fistulae (PAEF), which most commonly arise denovo from trauma or underlying neoplastic, infectious, or rheumatologic conditions. Although PAEF are associated with high mortality and morbidity, there have been several case reports documenting temporization of PAEF with stent grafts. These measures are effective in acute settings, often with unstable patients. This approach also provides time for the patient, family and their care team to decide on definitive treatment. This is especially important as many of these patients have serious underlying pathology, or will need to undergo extensive treatment for definitive care. Here, we present a case of PAEF in a patent with Stage IV renal malignancy. The bleed was controlled with the use of a 12 mm VIATORR endoprosthesis (W. L. Gore and Associates, Inc., Flagstaff, Arizona, USA), typically used in transjugular intrahepatic portosystemic shunts. The report demonstrates how an endoprosthesis originally designed for hepatic interventions can serve as an endovascular stent graft in extenuating circumstances. This report also aims to highlight the importance of following endovascular principles and the post-treatment care of patients with PAEF.
Keywords: Aortic repair, Endovascular intervention, Endovascular repair, Primary aortoenteric fistula, Transjugular intrahepatic portosystemic shunt graft

INTRODUCTION


Aortoenteric fistulae (AEF) are a rare cause of life-threatening gastrointestinal hemorrhage. These abnormal connections of the aorta and gastrointestinal tract are subdivided into two groups. Primary AEF (PAEF) occur without prior aortic intervention, while secondary AEF occur post-intervention. PAEF most commonly arise in the setting of infection, trauma, malignancy, or rheumatologic disorder; however, they remain exceedingly rare even in the setting of a known an aortic aneurysm.[1,2] Secondary AEF occur after the placement of an aortic graft and are relatively more common, but still rare, with prevalence estimates ranging from 0.36 to 1.6%.[3,4] There have been only a handful of documented cases of PAEF temporization with aortic stent grafts.[2,4-6] Here, we present a case of PAEF, temporized using a 12 mm VIATORR stent graft (W. L. Gore and Associates, Inc., Flagstaff, Arizona, USA) normally used during trans jugular intrahepatic portosystemic shunt creation.

CASE REPORT


The patient is a 44-year-old female with Stage IV renal malignancy. The patient initially presented with a suspected lower gastrointestinal bleed and severe hypotension. Computed tomography imaging of the abdomen demonstrated a large retroperitoneal mass is abutting the duodenum. There was contrast-extending form the aorta to the duodenum. These findings were concerning for an arterial-enteric fistula (Figure 1). The patient was emergently brought to IR for an angiogram.

Figure 1.jpg

Figure 1: 44-year-old female with aortoenteric fistula. Contrast enhanced arterial phase axial computed tomography demonstrates tumor encasing the inferior vena cava and aorta. There is contrast and air extending between the duodenum and aorta consistent with an aortoenteric fistula (white arrow).

Figure 1.jpg

Figure 1: 44-year-old female with aortoenteric fistula. Contrast enhanced arterial phase axial computed tomography demonstrates tumor encasing the inferior vena cava and aorta. There is contrast and air extending between the duodenum and aorta consistent with an aortoenteric fistula (white arrow).

Download as Power Point

After arterial access was obtained, a flush catheter was used to perform an infrarenal aortogram (Figure 2). This demonstrated active contrast extravasation from the infrarenal aorta into the second and third portions of the duodenum, consistent with a PAEF caused by the neoplastic invasion.

Figure 2.jpg

Figure 2: 44-year-old female with aortoenteric fistula. (a) Aortogram demonstrates extravasation of contrast from the aorta (white arrow). (b) Delayed image after aortogram demonstrates contrast pooling in the duodenum (white arrow).

Figure 2.jpg

Figure 2: 44-year-old female with aortoenteric fistula. (a) Aortogram demonstrates extravasation of contrast from the aorta (white arrow). (b) Delayed image after aortogram demonstrates contrast pooling in the duodenum (white arrow).

Download as Power Point

Given the degree of life-threatening hemorrhage, a decision was made to intervene. In this particular practice setting, no aortic stent grafts were available and the largest graft available was a 12 mm VIATORR endoprosthesis (W. L. Gore and Associates, Inc., Flagstaff, Arizona, USA) (Figure 3). Although not designed for placement in the aorta, the care team decided to use this graft as a last resort. The uncovered portion of the stent was placed at the level of the renal arteries. The covered portion was placed at the level of the AEF. A postplacement aortogram demonstrated a widely patent aorta, with adequate opacification of the renal arteries, and no further contrast extravasation into the duodenum (Figure 4). The patient was subsequently transferred back to the intensive care unit. After discussing goals of care, the patient and her family opted for no further intervention. The patient then passed away 48 h after intervention with comfort care measures.

Figure 3.jpg

Figure 3: (a) VIATORR graft in comparison to, (b) standard aortic stent EXCLUDER graft. Images are used with permissions from and provided courtesy of W. L. Gore and Associates.

Figure 3.jpg

Figure 3: (a) VIATORR graft in comparison to, (b) standard aortic stent EXCLUDER graft. Images are used with permissions from and provided courtesy of W. L. Gore and Associates.

Download as Power Point

Figure 4.jpg

Figure 4: 44-year-old female with aortoenteric fistula. Aortogram post graft (white arrow) placement demonstrates adequate control of the fistula with no further extravasation.

Figure 4.jpg

Figure 4: 44-year-old female with aortoenteric fistula. Aortogram post graft (white arrow) placement demonstrates adequate control of the fistula with no further extravasation.

Download as Power Point

DISCUSSION


Before the availability of endovascular therapy, PAEF were all treated surgically. In the setting of infected fields, surgeons were limited to using autologous or extra-anatomic grafts. These procedures had exceedingly high morbidity and mortality, due to acute complications such as hemorrhagic shock, sepsis, thrombus, and blowout.[7] Since then, there have been significant advances in endovascular therapy, but only a few cases of endograft repair of PAEF have been reported in the literature.[8]

This case report demonstrates that the off-label use of a stent graft can be used as a temporizing and life-prolonging measure in extenuating circumstances without significant pre-intervention planning. Despite the emergent nature of the case basic principles of endograft repair were followed - visceral arterial patency was maintained, and the defect was closed. Beyond these two principal concerns, there are several other factors to consider.

One important consideration is sizing, as the infrarenal aorta can vary in size greatly, from 12 to 18 mm in healthy women, but these can vary from 10 to 22 mm in otherwise healthy patients.[9] In the setting of pathology, aortic size can vary even further, affected by the invasive tumor, thrombus, atherosclerotic disease, increasing age, or with aneurysm. It is imperative that grafts are sized appropriately, with careful examination of pre-procedural imaging even in the acute setting. In addition to the graft diameter, it is important to denote the size of the fistula and proximity to visceral vessels, namely the renal arteries. The device chosen should maintain visceral patency while allowing for an adequate landing zone, in most endovascular aneurysm repairs about 1.5 cm is needed, less in some cases to prevent Type IA endoleaks. In cases of a juxta renal abnormality, one should consider a chimney graft or fenestrated graft, as in our case, to preserve renal patency. Distally the graft should ideally extend into the common iliac arteries to limit the possibility of Type IB endoleaks. Finally, one must also evaluate the number of lumbar arteries and the inferior mesenteric artery as these can contribute to a Type II leak.[10]

Review of imaging not only provides insight to size but can also help determine the best initial approach. Initial imaging can help denote the size of the iliofemoral arteries, in this case, a 10Fr sheath was needed, much smaller than most devices intended for aortic aneurysm repair. In addition, the vessel size, variant anatomy, calcifications, and tortuosity should be evaluated to avoid complications or to consider a vascular conduit.[10]

In this case, and others of neoplastic erosion, the patient requires close short-term follow-up. Although a graft may suffice to prevent subacute bleeds, the aorta and graft remain at risk of tumor erosion and infectious compromise.[11] In such cases future intervention should be carefully considered, taking account of the patient’s clinical status and goals of care.

CONCLUSION


AEF are rare entities associated with severe morbidity and mortality. Definitive and temporizing measures can be achieved with endovascular intervention. In extenuating circumstances, off-label use of stent grafts can be a life-saving measure. Even in the acute setting is important to review all available imaging and follow the principles of endovascular repair. After successful stenting, patients should be closely followed, as there are many possible complications.

REFERENCES


  1. , , , , . Endovascular techniques in the management of acute arterioenteric fistulas. J Endovasc Ther. ;11:89-93
    CrossRef    PubMed    
  2. , , , , . A case of primary aortoenteric fistula:Review of therapeutic challenges. Ann Vasc Surg. ;33(230):e5-13
    CrossRef    PubMed    
  3. , , , , , . Graft-related complications after abdominal aortic aneurysm repair:Reassurance from a 36-year population-based experience. J Vasc Surg. ;25:277-84
    PubMed    
  4. , , , , , . Aortoenteric fistulae temporization and treatment:Lessons learned from a multidisciplinary approach to 3 patients. Radiol Case Rep. ;12:331-4
    CrossRef    PubMed    
  5. , . Primary aortoenteric fistula:Two new case reports and a review of 44 previously reported cases. Eur J Vasc Endovasc Surg. ;12:5-10
    PubMed    
  6. , , , . Endograft for palliation of aortoenteric fistula. J Vasc Interv Radiol. ;28:1454
    CrossRef    PubMed    
  7. , , , . Impending aortoenteric hemorrhage:The effect of early recognition on improved outcome. Ann Surg. ;192:237-43
    PubMed    
  8. , , , . Endovascular management of acute bleeding arterioenteric fistulas. Cardiovasc Intervent Radiol. ;31:542-9
    CrossRef    PubMed    
  9. , , , . Infrarenal aortic diameter in the healthy person. Eur J Vasc Surg. ;8:89-95
    PubMed    
  10. , . Essentials of endovascular abdominal aortic aneurysm repair imaging:Preprocedural assessment. AJR Am J Roentgenol. ;203:W347-57
    CrossRef    PubMed    
  11. , , , , . Treatment of aortic graft infection in the endovascular era. Curr Infect Dis Rep. ;19:40
    CrossRef    PubMed    

How to cite this article:

MD AB, BS AS, DO SM. Partially Covered Transjugular Intrahepatic Portosystemic Shunt Endograft for Repair of a Primary Aortoenteric Fistula. Am J Interv Radiol 2018;2(18):1-4.



This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-Share Alike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.